Investigation of Optimal Solution Type and Volume for Inflation of Foley Catheter Balloon in Indwelling Urethral Catheterization

Objective: In this study investigates the effect of the type and volume of solution used in inflation of the foley catheter balloon on the development of urine leakage around the catheter during indwelling urethral catheterization. Material and Method: This randomized controlled experimental study was conducted with 128 patients hospitalized in the intensive care unit and carried out between 2015 and 2019. In each study group, different types and volumes of the solution including 10 mL 0.9% sodium chloride, 15 mL 0.9% sodium chloride, 10 mL sterile distilled water, and 15 mL sterile distilled water were used to inflate the foley balloon. Results: In the study, 7% of the patients who underwent indwelling urinary catheterization had urine leakage around the catheter. Urine leakage occurred on a mean of 11.33±8.22 days of catheterization. The urine leakage rate was calculated as 7.22/1000 catheterization days. Urine leakage rates of 10 mL 0.9% sodium chloride, 15 mL 0.9% sodium chloride, 10 mL sterile distilled water, and 15 mL sterile distilled water on the day of 1000 catheterization were 7.72, 5.60, 13.25, and 3.01 catheterization days, respectively. The most frequent urine leakage was in catheterizations using 10 mL sterile distilled water. Urine leakage was lower in catheterizations using 15 mL sterile distilled water compared to other groups. Conclusion: Using 15 mL of sterile distilled water to inflate the balloon can minimize urine leakage to ensure that the catheterization continues safely

Amaç: Bu çalışma foley kateter balonunun şişirilmesinde kullanılan solüsyon tipi ve hacminin kalıcı üretral kataterizasyon sırasında kateter etrafından idrar sızıntısı gelişimi üzerindeki etkisini incelemektedir. Foley balonun şişirilmesi için her bir çalışma grubunda 10 mL %0,9 sodyum klorür, 15 mL %0,9 sodyum klorür, 10 mL steril distile su ve 15 mL steril distile su olmak üzere farklı tip ve hacimLerde solüsyon kullanıldı. Gereç ve Yöntem: Bu randomize kontrollü deneysel çalışma, yoğun bakım ünitesinde yatan 128 hasta ile 2015-2019 yılları arasında yürütüldü. Bulgular: Çalışmada kalıcı üriner kateterizasyon uygulanan hastaların %7'sinde kateter çevresinde idrar sızıntısı gelişti. Kateterizasyonun ortalama 11,33±8,22 gününde idrar sızıntı meydana geldi. 1000 kateterizasyon gününde 10 mL %0.9 sodyum klorür, 15 mL %0,9 sodyum klorür, 10 mL steril distile su ve 15 mL steril distile su için idrar sızıntısı oranları sırasıyla 7,72, 5,60, 13,25 ve 3,01 gündü. En sık idrar sızıntı 10 mL steril distile suyun kullanıldığı kateterizasyonlarda gelişti. 15 mL steril distile su kullanılan kateterizasyonlarda idrar sızıntısı oranı diğer gruplara göre daha düşüktü. Sonuç: Balonu şişirmek için 15 mL steril distile su kullanmak, kateterizasyonun güvenli bir şekilde devam etmesini sağlamak için idrar sızıntısını en aza indirebilir.

Erişime Açık
Görüntülenme
138
08.07.2022 tarihinden bu yana
İndirme
1
08.07.2022 tarihinden bu yana
Son Erişim Tarihi
07 Aralık 2024 12:48
Google Kontrol
Tıklayınız
Tam Metin
Tam Metin İndirmek için tıklayın Ön izleme
Detaylı Görünüm
Eser Adı
(dc.title)
Investigation of Optimal Solution Type and Volume for Inflation of Foley Catheter Balloon in Indwelling Urethral Catheterization
Eser Sahibi
(dc.contributor.author)
AYŞE AKBIYIK
Yayın Tarihi
(dc.date.issued)
2022
Diğer Yazarlar
(dc.contributor.authors)
Ebru ÖZEL
Diğer Yazarlar
(dc.contributor.authors)
Perihan ÇETİN
Diğer Yazarlar
(dc.contributor.authors)
Leyla KHORSHID
Yayıncı
(dc.publisher)
İzmir Kâtip Çelebi Üniversitesi Sağlık Bilimleri Fakültesi
Tür
(dc.type)
Makale
Özet
(dc.description.abstract)
Objective: In this study investigates the effect of the type and volume of solution used in inflation of the foley catheter balloon on the development of urine leakage around the catheter during indwelling urethral catheterization. Material and Method: This randomized controlled experimental study was conducted with 128 patients hospitalized in the intensive care unit and carried out between 2015 and 2019. In each study group, different types and volumes of the solution including 10 mL 0.9% sodium chloride, 15 mL 0.9% sodium chloride, 10 mL sterile distilled water, and 15 mL sterile distilled water were used to inflate the foley balloon. Results: In the study, 7% of the patients who underwent indwelling urinary catheterization had urine leakage around the catheter. Urine leakage occurred on a mean of 11.33±8.22 days of catheterization. The urine leakage rate was calculated as 7.22/1000 catheterization days. Urine leakage rates of 10 mL 0.9% sodium chloride, 15 mL 0.9% sodium chloride, 10 mL sterile distilled water, and 15 mL sterile distilled water on the day of 1000 catheterization were 7.72, 5.60, 13.25, and 3.01 catheterization days, respectively. The most frequent urine leakage was in catheterizations using 10 mL sterile distilled water. Urine leakage was lower in catheterizations using 15 mL sterile distilled water compared to other groups. Conclusion: Using 15 mL of sterile distilled water to inflate the balloon can minimize urine leakage to ensure that the catheterization continues safely
Özet
(dc.description.abstract)
Amaç: Bu çalışma foley kateter balonunun şişirilmesinde kullanılan solüsyon tipi ve hacminin kalıcı üretral kataterizasyon sırasında kateter etrafından idrar sızıntısı gelişimi üzerindeki etkisini incelemektedir. Foley balonun şişirilmesi için her bir çalışma grubunda 10 mL %0,9 sodyum klorür, 15 mL %0,9 sodyum klorür, 10 mL steril distile su ve 15 mL steril distile su olmak üzere farklı tip ve hacimLerde solüsyon kullanıldı. Gereç ve Yöntem: Bu randomize kontrollü deneysel çalışma, yoğun bakım ünitesinde yatan 128 hasta ile 2015-2019 yılları arasında yürütüldü. Bulgular: Çalışmada kalıcı üriner kateterizasyon uygulanan hastaların %7'sinde kateter çevresinde idrar sızıntısı gelişti. Kateterizasyonun ortalama 11,33±8,22 gününde idrar sızıntı meydana geldi. 1000 kateterizasyon gününde 10 mL %0.9 sodyum klorür, 15 mL %0,9 sodyum klorür, 10 mL steril distile su ve 15 mL steril distile su için idrar sızıntısı oranları sırasıyla 7,72, 5,60, 13,25 ve 3,01 gündü. En sık idrar sızıntı 10 mL steril distile suyun kullanıldığı kateterizasyonlarda gelişti. 15 mL steril distile su kullanılan kateterizasyonlarda idrar sızıntısı oranı diğer gruplara göre daha düşüktü. Sonuç: Balonu şişirmek için 15 mL steril distile su kullanmak, kateterizasyonun güvenli bir şekilde devam etmesini sağlamak için idrar sızıntısını en aza indirebilir.
Kayıt Giriş Tarihi
(dc.date.accessioned)
08.07.2022
Açık Erişim Tarihi
(dc.date.available)
2022-07-08
Yayın Dili
(dc.language.iso)
eng
Konu Başlıkları
(dc.subject)
Indwelling urethral catheter
Konu Başlıkları
(dc.subject)
Saline
Konu Başlıkları
(dc.subject)
Balloon
Konu Başlıkları
(dc.subject)
Urine leakage
Konu Başlıkları
(dc.subject)
Sterile distilled water
Konu Başlıkları
(dc.subject)
Kalıcı üretral kateter
Konu Başlıkları
(dc.subject)
Salin
Konu Başlıkları
(dc.subject)
İdrar kaçağı
Konu Başlıkları
(dc.subject)
Steril distile su
Konu Başlıkları
(dc.subject)
Balon
Atıf için Künye
(dc.identifier.citation)
Akbıyık, A. , Özel, E. , Çetin, P. & Khorshid, L. (2022). Investigation of Optimal Solution Type and Volume for Inflation of Foley Catheter Balloon in Indwelling Urethral Catheterization . İzmir Katip Çelebi Üniversitesi Sağlık Bilimleri Fakültesi Dergisi , 7 (2) , 289-293 . Retrieved from https://dergipark.org.tr/tr/pub/ikcusbfd/issue/70141/957995
Alternatif Yayın Başlığı
(dc.title.alternative)
Kalıcı Üretral Kateterizasyonda Foley Kateter Balonunun Şişirilmesi İçin Optimal Solüsyon Tipi ve Hacminin İncelenmesi
ISSN
(dc.identifier.issn)
2458-9799
Yayının ilk sayfa sayısı
(dc.identifier.startpage)
289
Yayının son sayfa sayısı
(dc.identifier.endpage)
293
Dergi Adı
(dc.relation.journal)
İzmir Kâtip Çelebi Üniversitesi Sağlık Bilimleri Fakültesi Dergisi
Dergi Sayısı
(dc.identifier.issue)
2
Dergi Cilt
(dc.identifier.volume)
7
Haklar
(dc.rights)
Open access
Tek Biçim Adres
(dc.identifier.uri)
https://hdl.handle.net/11469/2083
Analizler
Yayın Görüntülenme
Yayın Görüntülenme
Erişilen ülkeler
Erişilen şehirler
6698 sayılı Kişisel Verilerin Korunması Kanunu kapsamında yükümlülüklerimiz ve çerez politikamız hakkında bilgi sahibi olmak için alttaki bağlantıyı kullanabilirsiniz.

creativecommons
Bu site altında yer alan tüm kaynaklar Creative Commons Alıntı-GayriTicari-Türetilemez 4.0 Uluslararası Lisansı ile lisanslanmıştır.
Platforms